A GAIN BOOSTED FOLDED CASCODE USING TELESCOPIC BOOSTING AMPLIFIERS WITH SWITCHED CAPACITOR INPUT LEVEL SHIFTERS

Bryce T. Bradford

Keywords

Biomedical Electronics, Electronic Medical Devices,Biosensors and transducers, Data and Signal Acquisition.

Abstract

This paper describes a single stage fully differential folded cascode operational transconductance amplifier (OTA) with fully differential gain boosting amplifiers designed in 1.2 VDC 130nm CMOS process technology. The amplifier is the driver for the analog to digital converter (ADC) of a blood pressure sensing medical implant [1] which uses wireless power transmission as its energy source [2]. The medical implant’s design specifications require an amplifier with wide output swing, high accuracy, and the bandwidth to drive a differential 5 pF load at 4 ksps with minimum power consumption. The gain boosting amplifiers are tele- scopic cascoded amplifiers, which, by their design, have an input and output common mode voltage relationship which is opposite to what is needed for a gain boosted cascode configuration. To correct for this, a method for DC shift- ing the amplifier’s input voltage using a switched capacitor network is presented.

Important Links:

Go Back